Study of interferences in the determination of lead by hydride generation-direct flame atomic-absorption spectrometry when oxidising agents are employed to increase the sensitivity

Juan R. Castillo, José M. Mir, Maria P. Colón, Carmen Martínez

The Analyst · 1985 · 23 citations · 0 references

Concepts

Abstract

The greatest sensitivity in the determination of lead by hydride generation-direct flame AAS is obtained in the presence of ammonium persulphate (0.012 µg). A complete interference study is reported, comparing the results obtained when hydrogen peroxide and ammonium persulphate are employed. The oxidation to a metastable tetravalent state of lead by the prior action of the oxidising agent is considered to be the mechanism by which the atomic-absorption signal in increased.