2014 · 20 citations · 10 references
EngineeringBiomedical EngineeringElectrical EngineeringAnalog ElectronicsBiomedical Signal ProcessingMicroelectronicsAnalog FiltersAnalog Integrated CircuitsCmos Circuit DesignDigital Signal ProcessingMixed-signal Integrated Circuit DesignMos CapacitorsAnalog-to-digital Converter DesignBiomedical ApplicationsCmos Analog Circuit DesignCmos Analog Integrated CircuitsComparator DesignDepletion-mode TransistorsSignal-to-noise RatioSupply Voltage Optimization
A 0.4-V opamp-less ΣΔ modulator for biomedical applications is presented. The continuous-time (CT) 2 <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">nd</sup> order ΣΔ modulator is realized using a passive filter topology. In order to obtain a small area, all capacitors are realized using parallel compensated depletion-mode MOSCAPs. To maximize the performance, genetic algorithms (GAs) are used to optimize the circuit for smaller area and higher signal-to-noise-plus-distortion ratio (SNDR). In the dynamic comparator, which is the only active block in the circuit, bulk-driven technique is used to obtain lower supply voltage operation. The circuit was designed in a 130 nm CMOS technology and, using a clock frequency of 2 MHz, has a bandwidth of 10 kHz to cover the most common bio-potential signals. The electrical transient noise simulation results show that the circuit achieves a peak SNDR of 58 dB and a dynamic range of 64 dB while dissipating only 410 nW, which corresponds to an energy efficiency of 32 fJ/conv.-step.
10