Emittance measurements at the ESRF

E. Tarazona, P. Elleaume

Review of Scientific Instruments · 1995 · 11 citations · 3 references

Concepts

Abstract

A new method of electron beam imaging is operating routinely at the ESRF. It uses the x-ray beam from an undulator. The white beam is monochromatized by a Si 400 Bragg reflection at 30 keV. The monochromatic beam is projected onto a fluorescent screen and imaged by a CCD camera. The emittance of the electron beam is deduced from the image size measured by the CCD camera and the independent knowledge of the beta function at the source point. This technique has measured emittances in the range of 0.2–20 nm rad with a precision of 20%. The principal limitation appears to be the single electron emission angular pattern which limits the minimum detectable emittance to 0.1 nm rad. This diagnostic has been routinely used for more than 1000 h. It is fully automated, resulting in a permanent display of the values of horizontal and vertical emittances without intervention by any operator.

References

3