Systematic and statistical errors in quantitative evaluation in TLC and HPTLC. 3. The determination of the primary statistical errors

19

Citations

2

References

1979

Year

Abstract

Abstract The paper describes the measurement and calculation of the primary statistical errors in the evaluation of TLC or HPTLC using in situ reflectance measurements. The primary errors are the errors of the sample volume spotted onto the plate, errors caused by the chromatographic process itself, the positioning error of the plate with respect to the centre of the light beam and the error in the light measurement. By scanning the same spot, the same track or the same plate under different conditions, the various errors can be determined.