MedicineHematologyPediatricsHemostasisPediatric HematologyPediatric TherapeuticsHemostatic DisordersPlatelet ActivationPlatelet FunctionPlatelet PhysiologyPrimary HemostasisVon Willebrand DiseaseBleeding Time AssessmentPediatric HemostasisPediatric Von Willebrand DiseasePlatelet Activation AssaysPlatelet Plug FormationVon Willebrand Disease Diagnosis
The PFA-100 system provides an in-vitro method of assessing primary platelet-related haemostasis by measuring the time (the closure time, or CT) taken for a platelet plug to occlude a microscopic aperture cut into a membrane coated with collagen and either epinephrine or ADP. We used the system to establish normal ranges for CTs in healthy children, adults and neonates. Mean CTs of healthy children were independent of the needle gauge used (21G or 23CG) for blood sampling; they were very similar to the mean CTs of healthy adults, but longer than mean CTs of healthy neonates. Although children with haemophilia had normal CTs, the PFA-100 system was found to be potentially useful in screening for von Willebrand disease in children.
5
Maturation of the hemostatic system during childhood
M Andrew, Patricia Vegh, Marilyn Johnston et al. · Blood · 1992 · 901 citations · Full text
Medicine, Hematology, Pediatrics +12
The bleeding time does not predict surgical bleeding [see comments]
Characterization of an In Vitro Platelet Function Analyzer, PFA-100™
Sourav K. Kundu, E Heilmann, Reynaldo Sio et al. · Clinical and Applied Thrombosis/Hemostasis · 1996 · 173 citations
Medicine, Hematology, Hemostasis +16
Evaluation of Platelet Function by PFA-100TM
Sourav K. Kundu, Reynaldo Sio, Angela Mitu et al. · Clinical Chemistry · 1994 · 30 citations