Optics Express · 2004 · 194 citations · 18 references
EngineeringBioengineeringBiomedical EngineeringBiomedical ImagingBiomedical OpticsMedical PhysicsLight-tissue InteractionOptical CoherenceOptical Coherence TomographyOptical TomographyScattering In TissueTissue Optical PropertiesAnisotropy FactorMultiple ScatteringMultiple Scattering EffectsOptical Characterization Of Biological TissueOptical Coherence Tomography ScatteringScatteringScattering Anisotropy FactorScattering Coefficient Estimation
We developed a new algorithm that fits optical coherence tomography (OCT) signals as a function of depth to a general theoretical OCT model which takes into account multiple scattering effects. With use of this algorithm, it was possible to extract both the scattering coefficient and anisotropy factor from a particular region of interest in an OCT image. The extraction algorithm was evaluated against measurements from an integrating sphere on a set of tissue phantoms and yielded valid results. Finally, a preliminary ex vivo OCT investigation on human aortic specimen indicated that the algorithm may contribute importantly to differentiation between normal and atherosclerotic arteries. We conclude that this algorithm may facilitate tissue characterization by OCT.
18
David Huang, Eric A. Swanson, Charles P. Lin et al. · Science · 1991 · 13.7K citations
Characterization of Human Atherosclerosis by Optical Coherence Tomography
Hiroshi Yabushita, Brett E. Bouma, Stuart L. Houser et al. · Circulation · 2002 · 1.2K citations
Medicine, Bioengineering, Cardiology +17
Ik‐Kyung Jang, Brett E. Bouma, Dong-Heon Kang et al. · Journal of the American College of Cardiology · 2002 · 1K citations