Review of Scientific Instruments · 2009 · 15 citations · 3 references
EngineeringElectrical EngineeringOpticsOptoelectronicsPhotonicsInfrared InstrumentationLaser OpticsLaser PhysicsOptical DetectorsOptical InstrumentationPhotodetectorsBeam AlignmentDetector SynchronizationDetectorsLaser AlignmentLaser Beam AlignmentOptical AlignmentPhotodiodesPosition-sensitive Detectors
A fully automatic alignment system for a pulsed infrared laser beam (5 ns pulses, 10 Hz repetition rate, 1.3 microm wavelength) was developed and tested. It compensates for long-term fluctuations of the beam initial position and direction-the automatic realignment is performed every 10 min, and lasts typically 1-2 min. The system adjusts the beam initial position with a maximum error of 0.5 mm (10% of the beam diameter) and the beam direction with a maximum error of 50 murad. The solution is based on two InGaAs quadrant photodiodes as the position detectors and two motorized mirrors controlled by a personal computer. The signals from the quadrant detectors are processed by a peak detector and digitized by an analog to digital converter, which is synchronized with the laser pulses.
3
Fast laser beam position control with submicroradian precision
S. Grafström, U. Harbarth, J. Kowalski et al. · Optics Communications · 1988 · 37 citations
Digital alignment system for a laser beam
E. Calloni, A. Brillet, C. N. Man et al. · Physics Letters A · 1994 · 10 citations