Environmental Science & Technology · 2009 · 28 citations · 19 references
Natural SciencesChemistryEnvironmental SciencesAnalytical ChemistrySensorsSpectroscopyGas SensorsInfrared DetectorsInfrared SpectroscopyLaser SpectroscopyNear-infrared SpectroscopyOptical SpectroscopyAmmonia DetectionAmmonia MonitoringCarbon DioxideGas Detection InstrumentsPhotoacoustic SpectroscopyPreconcentrationSample PreconcentrationSensor Intercomparison
This paper describes the first successful application of a preconcentration unitto increase the sensitivity of a photoacoustic ammonia concentration measuring instrument. A diode laser based near-infrared (1532 nm) photoacoustic ammonia monitoring instrument was combined with a tungsten (VI) oxide coated preconcentration unit to reach a sub-ppb detection limit with a compact, automatic measuring instrument. The system has no measurable cross-sensitivity to common atmospheric gases, most importantly to water vapor and carbon dioxide. The minimum detectable amount of ammonia is 2.9 ng, which means a minimum detectable concentration of 0.5 ppb with a 30-min measurementtime. Results of intercomparison measurements carried out both under laboratory and field conditions with reference to an electrochemical AMANDA instrument prove the applicability of the system in environmental ammonia concentration monitoring.
19
Method for determination of atmospheric ammonia
Martin Ferm · Atmospheric Environment (1967) · 1979 · 325 citations · Full text
Stéphane Schilt, Luc Thévenaz, Marc Niklès et al. · Spectrochimica Acta Part A Molecular and Biomolecular Spectroscopy · 2004 · 142 citations · Full text
Natural Sciences, Environmental Engineering, Environmental Sciences +15