Veterinary Surgery · 2007 · 39 citations · 33 references
MedicineOrthopaedic SurgeryOrthopedic SurgeryAnkle AnatomyBone AugmentationOrthopedic BiomechanicsBone BiomechanicsBone MechanicsExternal Skeletal FixationFixation BiomechanicsInternal FixationNail SurgeryBendingBending MechanicsDeformationIn Vitro BiomechanicsInterlocking Nail FixationInterlocking Nail SystemsLocked Plate FixationLocking
OBJECTIVE: To compare standard interlocking nails (ILN) with a newly designed ILN featuring an angle-stable locking mechanism (ILNn). STUDY DESIGN: Six experimental groups. SAMPLE POPULATION: Bone models (n=48) treated with 6 and 8 mm nails locked with screws or bolts (ILN6s, ILN8s, ILN6b, ILN8b, respectively), ILNn, and a 3.5 mm broad-DCP (br-DCP); n=4/testing mode. METHODS: Specimens were tested in torsion or 4-point bending. Construct compliance, deformation, and slack were statistically compared (P<.05). RESULTS: Regardless of testing mode, construct compliance was greater with smaller ILN. Screwed constructs were more compliant than bolted ones, with a significant difference between ILN6s and ILN6b in torsion. Plated constructs were significantly more compliant than the ILNn. Angular deformation was consistently greater with smaller ILN. Screwed ILN constructs sustained approximately 2 x the torsional deformation of the bolted ones (approximately 36 degrees [ILN6s] versus approximately 18 degrees [ILN6b]). Comparatively, ILNn constructs had significantly less torsional (approximately 8 degrees) and bending (approximately 4 degrees) deformation than other constructs. Whereas standard ILN constructs had slack in both modes, ILNn and br-DCP construct deformations consistently occurred without slack. CONCLUSIONS: Use of bolts rather than screws improved ILN mechanical behavior, but neither locking mechanism completely counteracted torsion and bending forces. Conversely, the ILNn angle-stable locking system eliminated torsional and bending slack, resulting in comparable mechanical performances between ILNn and plated constructs. CLINICAL RELEVANCE: The angle-stable locking mechanism of the new ILN eliminates all slack in the system; thus, interfragmentary motion will likely be reduced compared with standard ILN, which may improve the local environment for fracture healing.
33
Stephen R. Walsh, Rudy Reindl, Edward J. Harvey et al. · Clinical Biomechanics · 2006 · 125 citations