Review of Scientific Instruments · 1977 · 16 citations · 6 references
EngineeringElectrical EngineeringAnalog ElectronicsInstrumentationSensorsAnalog InstrumentationAnalog Integrated CircuitsDisplacement SensorsLow-noise Circuit DesignNoise In Analog CircuitsTransducer TechnologyDynamic RangeLock-in AmplificationMiniaturized TransducersPhase-sensitive DetectionPower ConsumptionTransducer CharacterizationTransducer DesignTransducer Drift
Differential capacitance displacement transducers (DCDTs) using audo frequency excitation and phase-sensitive detection are discussed, and their sensitivities are calculated. An inexpensive, low-power DCDT using a miniature lock-in amplifier made of integrated circuits is described. The lock-in amplifier has a digitally controlled phase shift between drive and reference signals, a full-scale output of ±10 V, and total noise and drift of about 1 mV peak to peak. The DCDT system has a detection limit of 0.01 nm/√H√z, a dynamic range between 60 and 80 dB, and a power consumption of less than 0.75 W. Parts cost for the system is about $200. A simplified DCDT not requiring the full lock-in amplifier is outlined, and improvements possible in the DCDT design are mentioned where appropriate.
6
The measurement and control of small displacements
R V Jones · Physics Bulletin · 1967 · 30 citations