Review of Scientific Instruments · 2010 · 103 citations · 8 references
Natural SciencesNuclear EngineeringPhysicsNuclear PhysicsRadiation MeasurementSpectroscopyGamma-ray DetectionGamma-ray DetectorsGamma-ray InstrumentationGamma-ray SpectroscopyInertial Confinement FusionNuclear FusionDensity MeasurementsDeuterium-tritium FusionGamma SpectroscopyGamma-ray Line EmissionGamma-ray SpectrometryImplosion DiagnosticsInelastic ScatteringInertial Confinement
The gamma reaction history (GRH) diagnostic is a multichannel, time-resolved, energy-thresholded γ-ray spectrometer that provides a high-bandwidth, direct-measurement of fusion reaction history in inertial confinement fusion implosion experiments. 16.75 MeV deuterium+tritium (DT) fusion γ-rays, with a branching ratio of the order of 10(-5)γ/(14 MeV n), are detected to determine fundamental burn parameters, such as nuclear bang time and burn width, critical to achieving ignition at the National Ignition Facility. During the tritium/hydrogen/deuterium ignition tuning campaign, an additional γ-ray line at 19.8 MeV, produced by hydrogen+tritium fusion with a branching ratio of unity, will increase the available γ-ray signal and may allow measurement of reacting fuel composition or ion temperature. Ablator areal density measurements with the GRH are also made possible by detection of 4.43 MeV γ-rays produced by inelastic scatter of DT fusion neutrons on (12)C nuclei in the ablating plastic capsule material.
8
ICF gamma-ray reaction history diagnostics
H. W. Herrmann, C. S. Young, J. M. Mack et al. · Journal of Physics Conference Series · 2010 · 49 citations · Full text
Natural Sciences, Nuclear Engineering, Nuclear Instrumentation +16
Observation of d–t fusion gamma rays (invited)
S. E. Caldwell, R. R. Berggren, Brent Davis et al. · Review of Scientific Instruments · 2003 · 36 citations