Concepedia

Publication | Closed Access

In Vitro Investigation of Cerebrospinal Fluid Leakage After Dural Puncture with Various Spinal Needles

64

Citations

8

References

1998

Year

Abstract

We compared several brands of pencil-point and standard cutting spinal needles of varying sizes. All pencil-point needles had less cerebrospinal fluid leakage, the least loss occurring with 26-gauge Atraucan needles. Electron microscopic examination of the dura after puncture showed characteristic findings with each needle type. We conclude that the combined cutting and pencil-point characteristics seen in the Atraucan needle may have clinical advantages.