Dielectrophoretic Sorting of Particles and Cells in a Microsystem

S. Fiedler, Stephen G. Shirley, Thomas Schnelle, Günter Führ

Analytical Chemistry · 1998 · 432 citations · 15 references

Concepts

TL;DR

Highly sensitive analytical techniques probe cellular events in very small volumes, yet effective sample handling and separation remain challenging; most devices rely on electrophoretic and chromatographic methods. The study demonstrates that negative dielectrophoresis forces generated by AC fields can be used for particle and cell sorting. The device employs a microchannel with platinum/titanium and ITO electrodes driven by high‑frequency AC, a laminar flow carrying particles, and a funnel‑aligner‑field cage‑switch assembly that concentrates, aligns, traps, and routes particles. The system handled micrometer‑sized latex particles and living mammalian cells, achieving concentration and switching at flow velocities up to ~10 mm s⁻¹, and its combination with high‑performance optical detection offers prospects for miniaturized flow cytometry.

Abstract

There are highly sensitive analytical techniques for probing cellular and molecular events in very small volumes. The development of microtools for effective sample handling and separation in such volumes remains a challenge. Most devices developed so far use electrophoretic and chromatographic separation methods. We show that forces generated by ac fields under conditions of negative dielectrophoresis (DEP) can also be used. Miniaturized electrode arrays are housed in a microchannel and driven with high-frequency ac. A laminar liquid flow carries particles past the electrodes. Modification of the ac drive changes the particle trajectories. We have handled latex particles of micrometer size and living mammalian cells in a device which consists of the following four elements: a planar funnel which concentrates particles from a 1-mm-wide stream to a beam of about 50-μm width, an aligner which narrows the beam further and acts to break up particle aggregates, a field cage which can be used to trap particles, and a switch which can direct particles into one of two output channels. The electrodes are made from platinum/titanium and indium tin oxide (ITO) on glass substrates. Particle concentration and switching could be achieved for linear flow velocities up to about 10 mm s-1. The combination of this new method with high-performance optical detection offers prospects for miniaturized flow cytometry.

References

15