IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control · 2010 · 36 citations · 40 references
Veterinary MedicineBioengineeringBiomedical EngineeringBiomedical ImagingDiagnostic ImagingUltrasonographyDiagnostic UltrasonographyElastographyUltrasonicsUltrasound ImagingUltrasound InstrumentationUltrasound PhysicsDoppler Ultrasound InstrumentationHigh-frequency Ultrasound SystemsPeriodic MotionPeriodic Motion GenerationPhase CorrectionUltrasonic ElastographyUltrasound Tissue CharacterizationUltrasound-based Tissue Characterization
In this article, a technique for measuring fast periodic motion is proposed. The sequencing used in this technique is similar to the one used in conventional color Doppler systems. However, a phase correction algorithm is introduced which compensates for the acquisition delays. Criteria for the types of motion which could be detected correctly by the system are developed and presented. Effective frame rates of several hundred hertz to a few kilohertz have been achieved with the system. Applications of the system in tissue elastography are presented together with experimental results from tissue mimicking phantoms.
40