EngineeringBiomedical EngineeringBiophysicsBiomedical ImagingBiomedical OpticsOpticsImagingLight-tissue InteractionMedical Imaging PhysicsNear-infrared ImagingOptical ImagingTissue OpticsDetector Response EstimationDetector Response FunctionsFull Width At Half MaximumTemporal GatingTime-of-flight Imaging
A new method for calculating the image resolution for a near-infrared time-of-flight tissue-imaging system is presented. The image resolution is calculated from the full width at half-maximum of the photon-path function along the midplane of the medium, integrated over all times of flight, and weighted by the time-resolved detector response. Detailed treatment of the optical gating mechanism shows that for some types of gating mechanism, there exists an optimal gating time beyond which the image resolution is not improved by arbitrarily decreasing the gating time. This theory predicts a limiting image resolution of~20% of the medium thickness, which is consistent with the research of others.