Publication | Closed Access
Test-Retest Reliability of Pure-Tone Thresholds from 0.5 to 16 kHz using Sennheiser HDA 200 and Etymotic Research ER-2 Earphones
116
Citations
13
References
2004
Year
Repeatability was similar for both transducer types. Intrasession test-retest repeatability from 0.5 to 12.5 kHz at each individual frequency including the frequency range susceptible to noise-induced hearing loss was excellent for both transducers. Repeatability was slightly, but significantly poorer in the frequency range from 14 to 16 kHz compared with the frequency ranges from 0.5 to 6 or 8 to 12.5 kHz. Measurements in the extended high-frequency range from 8 to 14 kHz, but not up to 16 kHz, may be recommended for monitoring purposes.