Biomechanical Assessment of Locked Plating for the Fixation of Patella Fractures

Simone Wurm, Peter Augat, V. Bühren

Journal of Orthopaedic Trauma · 2015 · 68 citations · 10 references

Concepts

Abstract

OBJECTIVE: To analyze the mechanical stability of locked plating in comparison with tension-band wiring for the fixation of fractures of the patella. METHODS: Biomechanical tests were performed on artificial foam patella specimens comparing an angular stable plate and monocortical screws with tension-band wiring. Tests were performed under combined tension and bending until failure simulating physiological loading of the tibia during walking. RESULTS: Tension-band wiring failed at 66% of the failure load of plating (1052 N, P = 0.002) and had 5 times larger fracture gap displacements (P = 0.002). CONCLUSIONS: Based on the biomechanical advantages, locked plating of the patella may constitute a reasonable alternative in the treatment of patella fractures.

References

10