Quantification of dose perturbations induced by external and internal accessories in ocular proton therapy and evaluation of their dosimetric impact
18
Citations
18
References
2013
Year
The degree of disturbance by external beam shaping devices remains low in ocular proton therapy and can be reduced by bringing accessories closer to the eye. Tantalum fiducial markers must be located in such a way that dose perturbation is not projected on the tumor. The treatment of patients with intraocular endotamponades must be carefully managed. It is fundamental that radiation oncologists and medical physicists are informed about the presence of such substances prior to the treatment.