The Journal of the Acoustical Society of America · 1963 · 39 citations · 0 references
Natural SciencesMaterials ScienceAcousticsInstrumentationUltrasoundAcoustic MeasurementPhysical AcousticsUltrasonicsUltrasound InstrumentationUltrasound PhysicsParallelism MeasurementQuartzStress WavesUltrasonic AcousticsUltrasonic AttenuationUltrasonic Materials CharacterizationUltrasonic Wave MeasurementUltrasonic Waves
Accurate ultrasonic attenuation and velocity measurements at megacycle frequencies require sample faces to be extremely parallel. To achieve ~10 % accuracy in quartz attenuation at 1000 Mc/s, the wedge angle must be ≤5 × 10⁻⁸ rad. Maintaining parallelism is critical, as echo patterns deteriorate when tolerances are violated; while it is barely achievable at 1000 Mc/s, at 25 000 Mc/s the required precision exceeds current instrument capabilities. These constraints apply only under rather special conditions.
The measurement of ultrasonic attenuation and velocity at megacycle frequencies demands that the sample faces be parallel within limits that require extreme care. The importance of maintaining the tolerances in question is shown in terms of the echo patterns for various cases in which the tolerances are held and in which they are not. In order to measure attenuation of crystalline quartz to within about 10% at 1000 Mc/sec, the required wedge angle θ should be about 5×10−8 rad or less. This turns out to be just barely possible. under rather special conditions. At the highest frequencies attained thus far (i.e., about 25 000 Mc/sec) with ultrasonic stress waves, the limits on parallelism are so severe that there are at present no instruments capable of making the necessary measurements to control the sample parallelism.