EngineeringNuclear EngineeringElectromagnetismNuclear InstrumentationSpectroscopyMagnetic SensorsNeutron DetectionNeutron ShieldingEnvironmental Noise MitigationMagnetic Field EffectsMagnetic Field StrengthNeutron Detector SystemsNeutron SpectrometryNuclear Magnetic MomentsProportionalityStabilization
The Surrounding Field Compensation (SFC) system described in this work is installed around the four-layer Mu-metal magnetic shield of the neutron electric dipole moment spectrometer located at the Paul Scherrer Institute. The SFC system reduces the DC component of the external magnetic field by a factor of about 20. Within a control volume of approximately 2.5 m × 2.5 m × 3 m, disturbances of the magnetic field are attenuated by factors of 5–50 at a bandwidth from 10−3 Hz up to 0.5 Hz, which corresponds to integration times longer than several hundreds of seconds and represent the important timescale for the neutron electric dipole moment measurement. These shielding factors apply to random environmental noise from arbitrary sources. This is achieved via a proportional-integral feedback stabilization system that includes a regularized pseudoinverse matrix of proportionality factors which correlates magnetic field changes at all sensor positions to current changes in the SFC coils.
26
Improved Experimental Limit on the Electric Dipole Moment of the Neutron
C.A. Baker, D. Doyle, P. Geltenbort et al. · Physical Review Letters · 2006 · 1.3K citations · Full text
Characterization of Frequency Stability
James A. Barnes, R. Andrew, L.S. Cutler et al. · IEEE Transactions on Instrumentation and Measurement · 1971 · 797 citations
Flavor physics of leptons and dipole moments
M. Raidal, A. van der Schaaf, I. I. Bigi et al. · The European Physical Journal C · 2008 · 377 citations · Full text