Review of Scientific Instruments · 1989 · 32 citations · 3 references
Natural SciencesBiochemistryBiophysicsAnalytical ChemistryMolecular BiophysicsStructural BiologyX-ray DetectorsX-ray SpectroscopyScatteringStopped-flow SpectroscopyTime-resolved Imaging
A stopped-flow apparatus was constructed for x-ray scattering study at subzero temperature. It can be operated over a wide temperature range down to −20 °C with highly viscous solution (20 cP) successfully. We have applied the stopped-flow x-ray scattering method to many biological reactions. In particular, the association of E. coli ribosomal subunits was detected at −10 °C which was too fast to be detected at room temperature. Dissociation of E. foetida hemoglobin was measured by the stopped-flow x-ray scattering method combined with a time-resolved imaging plate as a detector.