Review of Scientific Instruments · 1933 · 10 citations · 1 references
Natural SciencesEarth SciencesMineralogySpectroscopic AnalysisSpectroscopyAnalytical CalibrationCalibrationInfrared InstrumentationInfrared SpectroscopyOptical SpectroscopySpectral CalibrationCalibration Of Infrared InstrumentsDispersion AnalysisDispersion CharacterizationInfrared Detector CalibrationMountingSpectrometer CalibrationSpectroradiometric CalibrationWavelength Calibration
The dispersion data for rocksalt originally determined by Paschen are corrected to 25°C with the use of temperature coefficients derived from the measurements of Liebreich. Values of the angle of minimum deviation of a 60° prism and the prism table angles from the sodium D lines for a spectrometer with the Wadsworth mounting of a 60° prism and plane mirror are calculated for 25°C. Tables are given with which corrections may be easily made for temperatures other than 25°C and for prism angles between 59 and 61°.
1