A novel eddy current compensation scheme for pulsed gradient systems

R. E. Wysong, David P. Madio, I. J. Lowe

Magnetic Resonance in Medicine · 1994 · 32 citations · 5 references

Concepts

Abstract

We describe a modified pre-emphasis network that uses a number of fixed time constants, only whose amplitudes are adjustable. In comparison with variable time constant pre-emphasis networks we have used, the fixed time constant network is more effective and easier to set on the fly. We have further developed a method to set the fixed time constant network mathematically using a linear least squares technique.

References

5