Journal of Breath Research · 2010 · 35 citations · 24 references
PharmacyPharmacokineticsPhysiologyCompartmental ModelingPharmacokinetic ModelingSpectroscopyPulmonary PhysiologyRespiratory MonitoringRespiratory PhysiologyRespiratory System ModelingBreath Analysis MethodologyCavity Ring-down SpectroscopyExhaled Air SamplingHuman Body ModelsMulti-compartment DynamicsMulti-compartment EliminationPulmonary Elimination KineticsTime Resolution
We have measured the acetylene concentration in the exhaled breath of 40 volunteers (31 non-smokers, nine smokers) using near-infrared cavity ring-down spectroscopy. The acetylene levels were found to be the same as in ambient air for non-smokers, whereas elevated levels were observed for smokers. Real-time measurements with sub-second time resolution have been applied to measure the elimination kinetics of acetylene in breath after exposure to tobacco smoke. Three exponential time constants can be distinguished from the data and these can be used to define the residence times for different compartments, according to the multi-compartment model of the human body.
24
CW cavity ring down spectroscopy
D. Romanini, A. Kachanov, N. Sadeghi et al. · Chemical Physics Letters · 1997 · 531 citations · Full text
Breath Carbon Monoxide as an Indication of Smoking Habit
E. T. Middleton, Alyn H. Morice · CHEST Journal · 2000 · 420 citations
Effects of respiratory apparatus on breathing pattern
J. Askanazi, P. A. Silverberg, Robert J. Foster et al. · Journal of Applied Physiology · 1980 · 292 citations