British Journal of Applied Physics · 1954 · 29 citations · 0 references
Natural SciencesEarth SciencesRadiation MeasurementSensorsSpectroscopyInfrared Remote SensingRadiation InstrumentationSpectral CalibrationTemperature MeasurementTemperature MetrologyThermal Remote SensingBrightness Temperature MeasurementsColor Temperature MetersDirect MeasurementEmissivity EstimationOptical PyrometryOptical ThermometryThermal Infrared Observations
Expressions are derived for the errors of the two types and from them a chart, from which these errors can be calculated for any pyrometer for all possible values of emissivity. The limitations of both types for direct measurement of true temperature are discussed and the errors likely to result when true temperatures are calculated from apparent temperature from previous knowledge of emissivities.