Review of Scientific Instruments · 2006 · 15 citations · 14 references
Natural SciencesPhysicsOpticsPlasma PhysicsSpectroscopyLaser SpectroscopyRadiation SpectroscopyX-ray DetectionX-ray DetectorsX-ray OpticsX-ray SpectroscopyDetector Spatial ResolutionEnergy DensityHigh-resolution X-ray SpectroscopyMicrochannel Plate DetectorsOpacitySpectrometer DesignSpectrometer ResolutionX-ray Spectra
The spectroscopic investigation of high temperature laser produced plasmas in general, and x-ray opacity experiments in particular, often requires instruments with both a broad coverage of x-ray energies and high spectral, spatial, and temporal resolution. We analyze the design, model the response, and report the commissioning of a spectrometer using elliptical crystals in conjunction with a large format, gated microchannel plate detector. Measurements taken with this instrument at the Janus laser facilities demonstrate the designed spectral range of 0.24–5.8keV and spectral resolution E∕ΔE>400, resulting in two to three times more spectral data than achieved by previous spectrometer designs. The observed 100ps temporal resolution and 35μm spatial resolution are consistent with the requirements of high energy density opacity experiments.
14
Indirect X‐Ray Line‐Formation Processes in Iron L‐Shell Ions
M. F. Gu · The Astrophysical Journal · 2003 · 470 citations · Full text
Spectroscopic absorption measurements of an iron plasma
P. T. Springer, D. J. Fields, B. G. Wilson et al. · Physical Review Letters · 1992 · 150 citations