Publication | Closed Access
Fabrication and Electrochemical Characterization of Nanoporous Silicon Electrode for Amperometric Urea Biosensor
12
Citations
28
References
2012
Year
NanosensorsEngineeringNanoporous SiliconBiochemical SensorsUrea Sensitive ElectrodesBiomedical EngineeringChemistryBiosensorsModified Gold ElectrodeChemical EngineeringBiosensing SystemsAmperometric Urea BiosensorAnalytical ChemistryNanosensorElectrochemical CharacterizationCation SensingChemical SensorPorous SensorImplantable SensorNanoporous Silicon ElectrodeElectrochemistryBiomedical SensorsMicrofabricationBiomedical DiagnosticsElectroanalytical SensorWearable Biosensors
We describe a new type of biosensor that employs a modified gold electrode based on nanoporous silicon (NPSi) for the electrochemical detection of urea. Urease (Urs) was covalently immobilized onto an Au/NPSi electrode functionalized with 3-mercaptopropionic acid (3-MPA). Amperometric calibration curves for both NPSi and planar silicon (PLSi)-based urea sensitive electrodes were compared in the range of 0.3 to 4.5 mM urea concentrations. The Michaelis–Menten constant ( K m ) was determined using the amperometric method. The electrochemical active area ( A ea ) of the 3-MPA/Au/NPSi electrode was evaluated using cyclic voltammetry (CV) and the result was compared with the 3-MPA/Au/PLSi electrode. Measured sensitivity of the Urs/SAMs/Au/NPSi electrode is ca. 2.05 µA mM -1 cm -2 and that of the Urs/SAMs/Au/PLSi electrode is ca. 1.10 µA mM -1 cm -2 . About 1.8 times of sensitivity increase is obtained in the Au/NPSi electrode.
| Year | Citations | |
|---|---|---|
Page 1
Page 1