EngineeringGeneticsDna MicroarraysDna AnalysisAnalytical MicrosystemsGenomicsBiomedical EngineeringMobile MicrobeadsDna NanotechnologyBiosensing SystemsBiomedical DevicesMolecular DiagnosticsMicrofluidicsMicroarray Data AnalysisDna SequencingBiological NanomaterialsBiomedical AnalysisFunctional GenomicsBioinformaticsBiomedical DiagnosticsMicrofabricationHepatitis C ViralComputational BiologyBiotechnologyParallel Dna DetectionLab-on-a-chipBiomemsSystems BiologyMedicineBiomedical ApplicationsPolystyrene Microbeads
A microfluidic system has been designed and constructed by means of micromachining processes to integrate both microfluidic mixing of mobile microbeads and hydrodynamic microbead arraying capabilities on a single chip to simultaneously detect multiple bio-molecules. The prototype system has four parallel reaction chambers, which include microchannels of 18 × 50 µm2 cross-sectional area and a microfluidic mixing section of 22 cm length. Parallel detection of multiple DNA oligonucleotide sequences was achieved via molecular beacon probes immobilized on polystyrene microbeads of 16 µm diameter. Experimental results show quantitative detection of three distinct DNA oligonucleotide sequences from the Hepatitis C viral (HCV) genome with single base-pair mismatch specificity. Our dynamic bead-based microarray offers an effective microfluidic platform to increase parallelization of reactions and improve microbead handling for various biological applications, including bio-molecule detection, medical diagnostics and drug screening.