One step integration of gold coated sensors with OSTE polymer cartridges by low temperature dry bonding

Niklas Sandström, Reza Zandi Shafagh, Carl Fredrik Carlborg, T. Haraldsson, Göran Stemme, Wouter van der Wijngaart

2011 · 14 citations · 2 references

DOIFull text

Open access

Concepts

Abstract

We propose and demonstrate a novel one step method to integrate gold coated sensors with cartridges by dry covalent bonding. The cartridges are replica molded in an UV-curable off-stoichiometry thiol-ene (OSTE) polymer, featuring an excess of thiol functional groups that covalently bond to the surface of gold coated sensors upon contact. The method is demonstrated by the integration of a gold coated quartz crystal to a microfludic OSTE cartridge. The resulting bond interface is shown to be completely homogenous and void free and the package is tested successfully to a differential pressure of up to 2 bars. The performance of the biosensor chip is evaluated by measuring the unspecific binding of 0.5% albumin, resulting in a total frequency drop 205 Hz. This approach delivers a simple but rapid high quality integration aiming for the production of robust, low cost and disposable biosensor chips.

References