Publication | Closed Access
An Outbreak of <i>Pseudomonas aeruginosa</i> Related to Contaminated Urodynamic Equipment
26
Citations
8
References
1997
Year
UrologyPathogenic MicrobiologyMicrobial ContaminationAntibioticsAntimicrobial SusceptibilityMicrobial DiseaseHealthcare-associated InfectionDisinfection ProceduresImproper ProceduresHospital EpidemiologyMicrobiologyInfection ControlContaminated Urodynamic EquipmentPoint SourceMedicineBacterial PathogensClinical MicrobiologyAntimicrobial Resistance
Investigation of an outbreak of multiresistant Pseudomonas aeruginosa urinary tract infections within our hospital identified the urodynamic laboratory as the point source. Disinfection procedures were inadequate, and transducers designated for single-patient use were used repeatedly and subsequently became contaminated. These improper procedures were part of a cost-containment effort.