Optimization and evaluation of the signal intensity change in multisection oxygen-enhanced MR lung imaging

Ralf L�ffler, Christian J. M�ller, Michael Peller, H. Penzkofer, M. Deimling, Martin Schwaiblmair, J�rgen Scheidler, Maximilian Reiser

Magnetic Resonance in Medicine · 2000 · 104 citations · 16 references

DOIFull text

Open access

Concepts

Abstract

The behavior of the signal intensity in MRI of human lungs was investigated during inhalation of pure oxygen. Nine volunteers were examined, five using a breath-hold and four using a non-breath-hold technique. Four coronal slices were acquired in each volunteer using an inversion recovery turbo spin-echo sequence. The inversion time of the sequence was optimized for maximum contrast. Breathing of pure oxygen and room air was alternated in the volunteers. Breath-hold and non-breath-hold cases were compared. Breathing pure oxygen lead to a statistically significant signal intensity increase (up to 18%) compared to breathing room air. In addition, T(1) maps were acquired during breathing 100% oxygen and room air. Inhalation of pure oxygen reduced the mean T(1) time of the lungs from 1280 (+/-85) msec to 1224 (+/-139) msec without breath-hold and from 1219 (+/-176) to 1074 (+/-92) msec with breath-hold. Therefore, an optimized sequence and measurement protocol provided significant signal intensity changes utilizing 100% oxygen. Magn Reson Med 43:860-866, 2000.