Biomedical Optics Express · 2013 · 235 citations · 48 references
EngineeringBioengineeringBiomedical EngineeringBiomedical ImagingBiomedical OpticsDiagnostic ImagingCardiovascular ImagingCerebral Blood FlowFlow MeasurementHemodynamicsOptical Coherence TomographyVascular ImagingBlood FlowBlood Flow MeasurementOptical Coherence Tomography AngiographyPhase InformationRetinal Blood Flow ImagingTime ScalesVascular Flow ImagingVascular Flow Velocity Measurement
The split-spectrum amplitude-decorrelation angiography (SSADA) algorithm was recently developed as a method for imaging blood flow in the human retina without the use of phase information. In order to enable absolute blood velocity quantification, in vitro phantom experiments are performed to correlate the SSADA signal at multiple time scales with various preset velocities. A linear model relating SSADA measurements to absolute flow velocities is derived using the phantom data. The operating range for the linear model is discussed along with its implication for velocity quantification with SSADA in a clinical setting.
48
David Huang, Eric A. Swanson, Charles P. Lin et al. · Science · 1991 · 13.7K citations
Optical Coherence Tomography of the Human Retina
Michael R. Hee · Archives of Ophthalmology · 1995 · 1.6K citations