Smaller-than-Expected Sevoflurane Concentrations Using the Sevotec 5® Vaporizer at Low Fill States and High Fresh Gas Flows

Michael Seropian, Berklee Robins

Anesthesia & Analgesia · 2000 · 11 citations · 1 references

Concepts

Abstract

Smaller-than-expected concentrations of sevoflurane were delivered by Sevotec 5 vaporizers (Datex-Ohmeda, Madison, WI) at low-fill states, 8% dialed concentrations, and high fresh gas flows. Clinically, this would lead to prolonged induction times. The risk of complications from prolonged inhaled induction could be increased when using a vaporizer with these characteristics.