Review of Scientific Instruments · 1968 · 41 citations · 4 references
EngineeringMaterials EngineeringMaterials ScienceInstrumentationMaterials CharacterizationMetrologyAccurate Mass MeasurementAnalytical CalibrationCalibrationSensor CalibrationGravimetric MeasurementQuartzQuartz Crystal MicrobalanceQuartz Crystal Microbalance AnalysisQuartz Oscillator MetrologyResponse Function CalculationResponse Function ModelingSensitivity
A commercial quartz crystal microbalance has been calibrated via simultaneous comparison with a reference vacuum electrobalance. Aluminum, copper, and gold thermally deposited on 10 MHz AT-cut oscillators gave a composite linear calibration to a precision of ±0.2%. For a uniformly distributed deposition, the ratio K of the fractional change of mass Δm/m to the fractional change of frequency Δf/f is shown to be 0.991 with a probable error of 0.6% for a configuration having peripheral contact restraints of the oscillators. The response function δf/δm (sensitivity variation over the face of the crystal) was determined to exhibit asymmetries about the central axes.
4
<i>The Properties of Silica</i>. Robert B. Sosman
D. J. F. · The Journal of Geology · 1928 · 166 citations