<mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" display="inline"><mml:mi>K</mml:mi></mml:math>-Fluorescent-X-Ray Relative-Intensity Measurements
62
Citations
5
References
1971
Year
Math XmlnsSample Absorption-Fluorescent X RaysX-ray SpectroscopyHealth SciencesPhysicsMeasurementNatural SciencesSpectroscopyX-ray DiffractionX-ray TechnologyScofield CalculationsX-ray FluorescenceChemistryX-ray OpticAtomic Fluorescence SpectroscopyRadiologyX-ray Imaging
Measured values of the relative intensities of several $K$-fluorescent x rays are reported for 28 elements ranging from calcium through plutonium. Solid-state Si(Li) and Ge(Li) spectrometers and a Bragg-diffraction spectrometer were used in making the measurements. Fluorescence was excited in thin samples by higher-energy x rays. Corrections were made to the measured relative intensities for sample absorption, air absorption, and spectrometer efficiency. The accuracy of most of the measurements is \ifmmode\pm\else\textpm\fi{} 2 to 5%. The results are compared with those of other experimenters and with the Scofield calculations.