Comparison Study of the Rates of Manual Peripheral Blood Smear Review From 3 Automated Hematology Analyzers, Unicel DxH 800, ADVIA 2120i, and XE 2100, Using International Consensus Group Guidelines
27
Citations
7
References
2012
Year
False-negative rates were higher than the recommended levels. Among 3 automated hematologic analyzers, XE 2100 showed the highest rate of slide review. Because the present study clearly shows that the slide review rates have distinct characteristics among the studied analyzers, each individual laboratory should consider selecting the most appropriate analyzer according to clinical characteristics. Analyzers with high sensitivity may be advantageous in outpatient settings for screening patients, whereas analyzers with high specificity may be beneficial in inpatient settings for efficient patient care.