British Journal of Radiology · 1956 · 31 citations · 2 references
MedicineOphthalmologyRadiation DosimetryRadiation MeasurementRadiation SafetyAbsorbed DoseDosimetry InstrumentationEquivalent DoseRadiation Dose MeasurementBeta DosimetryBeta Radiation DosimetryDose Rate EstimationRadiation Monitoring InstrumentsSpherical CapsSurface Dose MeasurementSurface Radiation DoseTissue-equivalent Materials
β-ray emitting ophthalmic applicators, using strontium 90 foil, have been developed following earlier work with phosphorus 32. Details are given of these strontium 90 applicators, which have the form of spherical caps. A number of physical investigations of these applicators are described. Dose-rate variations across the surface were studied using a scintillation probe; the same instrument was used to find the variation of dose-rate with depth in tissue equivalent material, and to compare the dose-rate from the applicators with that from a plane sheet of foil. The method used to determine in absolute units the surface dose-rate from this plane foil, and hence from the applicators, is also described. Surface dose-rates for applicators at present in use are of the order of 100 rad/min.
2
Measurements with a Cerium 144 Beta-ray Applicator
J. L. Haybittle · British Journal of Radiology · 1953 · 22 citations
Radiotherapy of diseases of the cornea
M. Lederman · Journal of the Faculty of Radiologists · 1952 · 18 citations