Microwave plasma continuous emissions monitor for trace-metals in furnace exhaust

P. Woskov, D. Y. Rhee, P. Thomas, D.R. Cohn, J.E. Surma, Charles H. Titus

Review of Scientific Instruments · 1996 · 35 citations · 10 references

Concepts

Abstract

A microwave plasma continuous emissions monitor has been successfully demonstrated for sensitive (<1 ppb), real time measurements of trace metals in furnace exhaust. The instrument uses a robust, up to 1.5 kW, 2.45 GHz microwave plasma sustained in a portion of the undiluted furnace exhaust flow for atomic emission spectroscopy. The waveguide device is constructed of refractory materials compatible with high-temperature environments (≳500 °C) and is flange mountable into the inside of the furnace exhaust duct. Fused quartz fiber optics in close proximity to the plasma flame transmit the UV through visible emission (190–690 nm) to three spectrometers for simultaneous monitoring of several metals. This instrument has been used for continuous monitoring for a 49 h period with 0.5 s time resolution on a dc graphite electrode arc furnace during a soil vitrification test. Results are presented for chromium, manganese, and iron emissions during soil loading operations.