Journal of the American Society for Mass Spectrometry · 2011 · 28 citations · 84 references
We describe the construction and application of a new MALDI source for FT-ICR mass spectrometry imaging. The source includes a translational X-Y positioning stage with a 10×10 cm range of motion for analysis of large sample areas, a quadrupole for mass selection, and an external octopole ion trap with electrodes for the application of an axial potential gradient for controlled ion ejection. An off-line LC MALDI MS/MS run demonstrates the utility of the new source for data- and position-dependent experiments. A FT-ICR MS imaging experiment of a coronal rat brain section yields ∼200 unique peaks from m/z 400-1100 with corresponding mass-selected images. Mass spectra from every pixel are internally calibrated with respect to polymer calibrants collected from an adjacent slide.
84
Liam A. McDonnell, Ron M. A. Heeren · Mass Spectrometry Reviews · 2007 · 1.1K citations
J.W. Gauthier, T. R. Trautman, D. B. Jacobson · Analytica Chimica Acta · 1991 · 644 citations
Collision-activated Dissociation, Natural Sciences, Spectroscopy +8