Analytical Chemistry · 1996 · 11 citations · 7 references
Natural SciencesMaterials EngineeringMaterials ScienceBuilding MaterialsElemental AnalysisMaterials CharacterizationGypsum MaterialsPowder ProcessingX-ray DiffractionCalibration StandardsCrystallite Size AnalysisCrystallite Size DeterminationMillingPowder DiffractionPowder Diffraction AnalysisPowder X-ray DiffractionX-ray Diffraction Analysis
An X-ray diffraction procedure for quantitative chrysotile asbestos analysis in a gypsum-based matrix is discussed. The procedure includes calibration standards preparation, an EDTA treatment to remove gypsum, sample milling, and X-ray diffraction analysis. The effect of each step is discussed. Morphological and crystal structural changes due to the sample preparation procedure were also characterized using techniques such as SEM, IR, NMR, and ICP. Through a statistically designed experiment, it is demonstrated that X-ray diffraction can be used as a reliable technique in gypsum-based matrices to determine chrysotile concentration from 0.25% to 2.5% with a standard deviation of 0.07%.
7