Temperature dependence of gauge factor and magnetoresistance of some platinum-tungsten strain gauges

D. I. Bower

Journal of Physics E Scientific Instruments · 1972 · 10 citations · 3 references

Concepts

Abstract

The gauge factor of some Denfoil type 1831 NE foil gauges was found to increase monotonically by 5.7% of its room temperature value of 4.45 as the temperature was lowered to 4K. The magnetoresistance is negligible at room temperature and even at 4K and 1T is only equivalent to a strain of about 1.6*10-6, with an anisotropic part of about 0.5*10-6.

References

3