PharmacyChemistryImmunologyAnalytical ChemistryMass Spectrometric AnalysisProteomicsCharged Particle DetectionDetector SystemsMass AnalyzersMass SpectrometryPeptide AnalysisCharge State DistributionsCharge-state DistributionMatrix-assisted Laser Desorption Ionization Time-of-flight Mass SpectrometryMatrix-assisted Laser Desorption/ionization Time-of-flight Mass SpectrometryNew Instrument DesignProtein Mass SpectrometrySensitivityTime-of-flight Mass Spectrometry
Presented are initial results from the first commercially available matrix-assisted laser desorption/ionization time-of-flight mass spectrometer specifically designed for the sensitive detection of very high mass ions (macromizer, Comet AG). This new instrument utilizes a 16-element superconducting tunnel junction detector coupled with a fully adjustable gimbal-mounted ion source/focusing region that allows unparalleled sensitivity for detection of singly charged high molecular weight ions. Using this new technology, the singly charged ions in the megadalton region are detected from immunoglobulin M and von Willebrand factor proteins. This detector technology also measures the kinetic energy of the particles impacting the detector, which can be correlated to the charge of the particles. Immunoglobulin G and streptavidin were used to demonstrate the ability of the macromizer instrument to detect high-mass ions and to discern the charge state of the ions.