Natural SciencesMechanical EngineeringMechatronics EngineeringActuatorsMicroelectromechanical SystemsOpticsConfocal MicroscopyFluorescence MicroscopyLight MicroscopyMicroscopic ExaminationOptical InstrumentsScanning Probe MicroscopyElectromagnetic ActuatorsFeedback DesignMicroscope ImagingMicroscopic ImagingMicroscopic ObservationMicroscopyMicroscopy ImagingPiezoelectric Sensors
The system is especially suitable for high‑resolution confocal UV microscopy. The authors designed and constructed a high‑speed mechanical scanning stage for microscopic applications. The stage uses a feedback design with electromagnetic actuators, piezoelectric sensors, and motor‑driven screws for coarse adjustment, and incorporates an exceptionally stable optical system. The stage achieves scanning of up to 1 mm² areas at 300 lines/s with positional resolution better than 0.01 µm.
A high-speed mechanical scanning stage for microscopic applications has been designed and constructed. It is especially suitable for high-resolution confocal UV microscopy. It is a feedback design using electromagnetic actuators and piezoelectric sensors with motor-driven screws for coarse adjustment. Scanning of areas up to 1 mm square at up to 300 lines per second is possible with positional resolution of better than 0.01 μm. An exceptionally stable optical system is also described.