Journal of High Resolution Chromatography · 1986 · 21 citations · 10 references
Natural SciencesChemistryAnalytical ChemistrySpectroscopic AnalysisSpectroscopyEmission SpectroscopyGas ChromatographyMicrowave SpectroscopyOptical SpectroscopyPlasma SourcesUltraviolet SpectroscopyAtmospheric PressureCapillary Gas ChromatographyExcitationInstrumental Background CorrectionMicrowave PlasmasPhotodiode ArraysPlasma ChemistryPlasma Emission Spectroscopy
Abstract A microwave‐induced plasma emission spectrophotometric detector (MIPD) was used as an element‐specific detector for capillary column gas chromatography. The atmospheric pressure microwave helium plasma generated with an original device called a SURFATRON was used as an atomization and excitation source. Combining a photodiode array spectrophotometer with the above system made the emission spectrophotometric detector very powerful. A wide range of spectra could be instantly monitored without any mechanical device. However, the spectrum of atmospheric helium emission plasma was complicated by the presence of air around the plasma discharge. An on‐line background correction scheme was developed to handle such complicated spectra.
10