Embolization of Hydrophilic Catheter Coating to the Lungs

Robert W. Allan, Hassan Alnuaimat, William D. Edwards, Henry D. Tazelaar

American Journal of Clinical Pathology · 2009 · 58 citations · 4 references

Concepts

Abstract

Hydrophilic coatings are used on intravascular devices to facilitate ease of manipulation and to minimize complications such as thrombosis during a procedure and vasospasm after a procedure. We report a case of embolization of hydrophilic coating of a central venous catheter to the lung that resulted in cavitary lung nodules in a 34-year-old woman. The microscopic features of this unusual complication warrant emphasis so that pathologists will not overlook the embolic foreign material and make a faulty diagnosis of noniatrogenic granulomatous vasculitis.

References

4