MedicineOncologyRadiation OncologyClinical Radiation OncologyRadiation DosimetryTreatment PlanningDose CalculationPatient-specific DosimetryRadiation Dose MeasurementRadiation TherapyRadiation Treatment PlanningTreatment Planning Dose CalculationDose PlanningDose StatisticsRadiation Dose VerificationRadiotherapy Plan EvaluationTreatment Plan EvaluationTreatment Planning DosimetryTreatment Planning System Dose VerificationTreatment Planning System Verification
A comparison of isocenter dose calculated by a commercial intensity modulated radiation therapy treatment planning system and independent monitor unit verification calculation (MUVC) software was made. The percent disparity between the treatment plan and MUVC doses were calculated for 507 treatments (head and neck, prostate, abdomen, female pelvis, rectum and anus, and miscellaneous) from 303 patients. The MUVC calculated dose was, on average, 1.4% higher than the treatment planning dose, with a 1.2% standard deviation. The distribution of the disparities appeared to be Gaussian in shape with some variation by treatment site. Based on our analysis, disparities outside the range of +/-3% about the mean value should be checked and resolved prior to treatment delivery.
8
Dosimetric verification of a commercial inverse treatment planning system
Lei Xing, Bruce Curran, R Hill et al. · Physics in Medicine and Biology · 1999 · 108 citations