Physics in Medicine and Biology · 2000 · 79 citations · 7 references
MedicineOncologyRadiation OncologyRadiation DosimetryRadiation MeasurementTreatment PlanningIntensity-modulated RadiotherapyPhoton DosimetryRadiation TherapyRadiation Treatment PlanningTreatment Planning Dose CalculationTreatment Planning PhysicsIntensity Modulated Radiation Therapy Dose CalculationIntensity ModulationIntensity-modulated Radiation Therapy DosimetryModulated Photon TherapyPhoton Dose CalculationRadiation Dose EstimationRadiotherapy Dose PlanningTreatment Planning System Dose Verification
An important issue in intensity modulated radiation therapy (IMRT) is the verification of the monitor unit (MU) calculation of the planning system using an independent procedure. Because of the intensity modulation and the dynamic nature of the delivery process, the problem becomes much more involved than that in conventional radiation therapy. In this work, a closed formula for MU calculation is derived. The approach is independent of the specific form of leaf sequence algorithms. It is straightforward to implement the procedure using a simple computer program. The approach is illustrated by a simplified example and is demonstrated by a few CORVUS (NOMOS Corporation, Sewickley, PA) treatment plans. The results indicate that it is robust and suitable for IMRT MU verification.
7
The Physics of Radiation Therapy
JAMA · 1953 · 1.2K citations
Dosimetric verification of a commercial inverse treatment planning system
Lei Xing, Bruce Curran, R Hill et al. · Physics in Medicine and Biology · 1999 · 108 citations