Journal of Physics Conference Series · 2010 · 16 citations · 1 references
EngineeringComputer EngineeringElectrical EngineeringAnalog ElectronicsInstrumentationMetrologyAnalog InstrumentationData Acquisition SystemsDigital InstrumentationDigital Signal ProcessingFpga ArchitectureSensor InterfacingDemodulationDigital Measurement SystemsLocal Data ProcessingMeasurement Noise ReductionPhase NoiseProcessingSignal AcquisitionSignals
A novel MIT Direct Digitising Signal Measurement (DDSM) module has been developed aiming to replace the centralised NI PXI system and PC processing of the Cardiff Mk2 MIT system, thus offering potentially faster measurement cycles. The proposed module replaces the signal acquisition and offers local processing. The core of the system is a Xilinx Spartan-3 FPGA, paired with a dual 14-bit ADC capable of 120 MS/s. The FPGA provides a flexible and fast platform for data acquisition and processing. The phase is measured using a two channel phase sensitive detection via I/Q demodulation. Built-in averaging reduces the data to a single signal period of 12 samples before multiplying and accumulating the data with the I/Q signals. The system provides the I/Q values for both channels directly, eliminating the long download and processing times required by the centralised NI PXI system currently used in the Cardiff Mk2 MIT system. The module acquires both the measurement and the reference signal and has a phase noise as low as 7.5m° using a measurement time constant of 6ms. The phase drift over 6 hours is considerable at 119m°. Details of the module, circuits and algorithms employed are provided as are the results of the performance measurements.
1