Optics Letters · 2012 · 62 citations · 12 references
EngineeringBiomedical EngineeringBiophysicsBiomedical ImagingBiomedical OpticsBiophotonicsLight-tissue InteractionOptical ImagingPhotoacoustic ImagingScattering In TissueTissue Optical PropertiesTissue OpticsDepth Penetration In Photoacoustic ImagingOptical Scattering In TissuePenetration DepthPhotoacoustic Imaging InstrumentationPhotoacoustic Imaging SystemsPhotoacoustic MicroscopyReflection EffectSpatial Resolution In Photoacoustic Imaging
Recent developments of optical-resolution photoacoustic microscopy (OR-PAM) have improved its spatial resolution and imaging speed. However, the penetration depth of OR-PAM is still limited to ~1 mm in tissue, owing to the strong tissue scattering. Here, we have developed double-illumination PAM (DI-PAM), which illuminates the sample from both top and bottom sides simultaneously. Through phantom and in vivo experiments, we have demonstrated for thin targets that DI-PAM has a penetration depth of ~2 mm in tissue at 532 nm and a focal zone of 260 μm, both significant improvements over traditional reflection or transmission-mode OR-PAM.
12
Label-free oxygen-metabolic photoacoustic microscopy in vivo
Junjie Yao, Konstantin Maslov, Yu Shrike Zhang et al. · Journal of Biomedical Optics · 2011 · 329 citations · Full text