Publication | Open Access
Complex wavefront shaping for optimal depth-selective focusing in optical coherence tomography
113
Citations
34
References
2013
Year
EngineeringMicroscopyCoherenceBiomedical EngineeringTissue ImagingPenetration DepthRadiologyHealth SciencesPhotonicsMedical ImagingOphthalmologyBiophotonicsComputational Optical ImagingOptical ImagingBiomedical ImagingOptimal Depth-selective FocusingMultiple Light ScatteringOptical Coherence TomographyComplex WavefrontTomography
We report on an approach to exploit multiple light scattering by shaping the incident wavefront in optical coherence tomography (OCT). Most of the reflected signal from biological tissue consists of multiply scattered light, which is regarded as noise in OCT. A digital mirror device (DMD) is utilized to shape the incident wavefront such that the maximal energy is focused at a specific depth in a highly scattering sample using a coherence-gated reflectance signal as feedback. The proof-of-concept experiment demonstrates that this approach enhances depth-selective focusing in the presence of optical inhomogeneity, and thus extends the penetration depth in spectral domain-OCT (SD-OCT).
| Year | Citations | |
|---|---|---|
Page 1
Page 1